"Classic"
make Photo Stability chamber is design to
suit standards for photogenic effect along
with temperature & Humidity for not
less than 1.2 million Lux hours and ultra
violet source of energy for not less than
200 watts hours/sq. meter.
As per ICH & cGMP guidelines.
Range
: Temperature
range
: 10°C to 40°C
Temperature accuracy :
+/- 0.5°C
Temperature Uniformity :
+/- 1°C
Humidity range : 40
% r.h. to 80% r.h.
Humidity Accuracy : +/-
2% r.h.
Humidity Uniformity :
+/- 3% r.h.
Application
: PHOTO STABILITY CHAMBER are used for
studies related to Photogenic effect on
the samples exposed to U.V. & day light
illuminations. It is design to create Nature
Climatic conditions with the help of Dry
air, refrigeration & Humidity system
built-in. This Chambers are basically in
application with Pharmaceuticals, &
also Electro-mechanical R&D test to
confirm the successful result of the product
under test.
Construction :
Photo Stability Chambers are constructed
with highly Insulated "PUF" or Glass wool insulation.
The body is either galvanized or
stainless steel 304 quality, where as inner
is 304 S.S. Door flush
type with full size glass door for viewing
inside objects with silicon gasket &
locking arrangement.
Documentation :
Provided with DQ,IQ,OQ,PQ protocol &
Validation report.
Models
CS-90a
CS-90b
CS-90c
CS-90d
Chamber
Volume
90.0
ltrs.
170.0
ltrs.
285.0
ltrs.
340.0
ltrs.
Cubic
Feet
3
6
10
12
Internal
Size cms.
40
X 40 X 57
50
X 40 X 85
57
X 57 X 88
60
X 60 X 95
External
Size cms.
53
X 68 X 135
63
X 68 X 163
70
X 85 X 166
73
X 88 X 173
Shelves
2
2
3
3
Range
+
10 C upto + 40 C
Controller
Accuracy
+/-0.1
C of set Value for Temp / + /- 2 % (Humidity)
Uniformity
+/-
0.5 C for Temp. +/- 3 % for Humidity
Photogenic
effect
Ultra
Violet wavelength 365nm. (measured
in hours/square meter)
Florescent light 4 nos. of >200
watts (measured in Lux)
Heat
up Time
30
min. upto 40 C without load.
Cool
Down time
15
min. upto + 10 C without load
Cooler
Plate
Fin Cooler
Compressor
Hermetically
Sealed Brand New Kirloskar Compressor
Controller
Microprocessor
based programmable P.I.D controller.
Sensor
Pt-100
for Dry & Wet Bulb / Capacitor
type for Humidity.
Internal
Chamber
Stainless
Steel 304 / 316L quality.
External
Stainless Steel
Matt Finish / Mild
Steel Powder Coated
Insulation
Body
Puf
50mm / Glass Wool 65mm
Door
Thermo
Cole.
Inner
Glass Door
Inner
Full Size Sea through Acrylic 8 mm thick.
Forced
Air Draft
7"
Tangent Flow by Aluminium blower
Air
Circulation Motor
1/12
Hp, 1440 R.P.M.
Steam
Injection System
I.S.I.
Mark Kettle Heater / ultrasonic humidifier.
Water
tank Cartridge
Reservoir
4 ltrs
4 ltrs
9 ltrs
9 ltrs
Boiler
4 ltrs
4 ltrs
9 ltrs
9 ltrs
Water
quality
pH6.2
to 7.2
Safety
Device
Water
Level alarm, over heating or humidity
soft alarm.
Optional
Imported
Sensor
0
to 100% R.H., Accuracy of +/-2%
Data
Acquisition
21
CFR part 11 compliance.
Computer
Interface
RS
485 / RS 232 interface for multiple
/ single communication port.
Safety
device
Thermostat
or Energy Regulator
Cyclic
Timer
0
to 24 hrs X 7 days cyclic ON / OFF timer
for illuminating condition.
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Back | HOME | Introduction
| Co. Profile
| Products | Clients | Feedback
| Contact
| Next TOP